Clean Environment and Microscope (Phase Contrast & Fluorescence)in JEM

Overview

The Clean Bench consists of two compartments; the Disinfection Chamber (DC) and the Operation Chamber (OC).
Air circulated inside is kept clean by HEPA filters.
Crew members operate the experiment materials with gloves from outside to prevent the ambient air contaminating.

 

Microscope

Clean Bench has a specially designed microscope in the OC.
Bright-Field, Phase-Contrast and Fluorescence Microscope modes are all available. Objective Lens can be switched among four magnification levels such as x4, x10, x20, x40.

 

Tele-science

Remote Controlled from ground by telemetry command

The microscope image and operation scene in the OC can be monitored simultaneously on the ground.
The microscope X,Y stage and microscope focus can be controlled from the ground.
The microscope mode (for example, from Phase Contrast to Fluorescence) and Fluorescence filter can be altered from the ground by telemetry command.

 

Other Functions

Operation in OC

The microscope image and operation scene in the OC can be monitored simultaneously on the ground.
The microscope X,Y stage and microscope focus can be controlled from the ground.
The microscope mode (for example, from Phase Contrast to Fluorescence) and Fluorescence filter can be altered from the ground by telemetry command.

 

SPEC

No. Item Specification
1 Operation Chamber (OC)
Outline : sealed container with 3 gloves for operation
Sterilization : Ultra Violet Lamp
Volume : 62 liters
2 Disinfection Chamber (DC)
Outline : Sealed container with 2 gloves for operation
Volume : 16 liters
3 Environmental Control
Gas Monitor system
  Toxic gas alarms in the OC
    Ethanol, Chloroform, Ammonia, Glutaraldehyde, Acetone, Acetic Acid, Formaldehyde, Methanol.
Environmental Control
  Thermal Control 20 to 38±2 deg. C
Air dust filtering system
  HEPA filter x2
4 User Support Function
Bright Field, Phase Contrast and Fluorescence Microscope
  Image captured by CCD Camera and displayed on CB LCD Monitor
Objective Lens:
  x4, x10, x20, x40 (for Bright Field, Phase Contrast Obsevation) x40 (for Fluorescence Obsevation)
Focus:
  Vertical
Horizontal
0 - 10 mm (Z position)
±12.5 mm each (X,Y position)
Lamp:
  Halogen Lamp 12 V, 50 W (for Bright Field, Phase Contrast Observation)
Xenon Lamp 125 W (for Fluorescence Observation)
Internal OC Camera
Operation Monitoring in OC
Joy-Stick
Microscope X-Y Stage Position and Focus Position Control
OC Switch Box
Microscope Control in OC
LCD monitor
Diplay Image of Microscope, Monitor Camera in OC, user Camera
5 User Interface
(Experiment Equipment I/F)
Connector for user camera
  NTSC: 1 line
Power Resource Connector
  5VDC: 1 A, +12 VDC: 1.25 A, ±15 VDC: 0.17 A
 